800 Dosino - General
The 800 Dosino is a flexible dosing drive which, as an MSB device, can be used with a host of Metrohm main devices. Together with the intelligent 807 Dosing unit, it is well suited for both simple dosing and for automation and liquid handling tasks, such as sample transfer or pipetting.
Note MagIC Net only supports the use of intelligent 807 Dosing units!
Configuration MSB Device
An MSB device is not entered in the device table. In the configuration it is only listed under the MSB connections of the main device to which it is connected (e.g. 850 Professional IC or 858 Professional Sample Processor).
800 Dosino - Method
An 800 Dosino (as is the case with all MSB devices) cannot be added separately to a method. It is added to the list of devices by adding the main device to which it is connected to the method.
If the method is created without connected devices, the MSB device can be added separately. In this case, the MSB connector on the main device must be entered manually.
In the device group, the 800 Dosino behaves like a stand-alone device. It can be selected and displayed graphically in the upper area of the device window.
800 Dosino - Start parameters
Tab: Method ? Devices ? Dosino Solution
Select the solution for the Dosino that should be available when the hardware is started with [Start HW]. A check is carried out in the method run to ensure that the correct solution has been put on the dosing device. When the method is started, the working life and GLP test interval are checked for the selected solution.
Selection not defined | Solutions from the solution table Default value not defined not defined
The solution is not checked.
Solutions from the solution table
Only \solutions, i.e. solutions that are assigned to an 807 Dosing unit, can be selected. Connector
Shows the main device and the MSB connector which the 800 Dosino is on.
800 Dosino - Commands
Overview of the time program commands available for the 800 Dosino:
Dosing
Dialog window: Method ? Time program ? Edit ? 'Device' - Dosing
The device-dependent time program command Dosing (command with feedback) doses the specified volume via the defined port. There is no automatic filling beforehand or afterwards.
Parameters
Port
Port via which dosing takes place. Formula input possible.
Range Default value Volume
1 ... 4 1
Volume to be transported. Formula input possible.
Range Default value Dosing rate
0.0000 ... 99999.9000 mL 1.0000 mL
Speed at which discharge takes place. The maximum rate depends on the cylinder volume of the buret used. Formula input possible.
Cylinder volume 2 mL
Range Default value Selection Range Default value Selection Range Default value Selection Range Default value Selection Range Default value Selection
0.01 ... 6.00 mL/min 1.00 mL/min maximum
Cylinder volume 5 mL
0.01 ... 16.00 mL/min 1.00 mL/min maximum
Cylinder volume 10 mL
0.01 ... 33.00 mL/min 1.00 mL/min maximum
Cylinder volume 20 mL
0.01 ... 66.00 mL/min 1.00 mL/min maximum
Cylinder volume 50 mL
0.01 ... 166.00 mL/min 1.00 mL/min maximum
Filling rate
Speed at which filling takes place. The maximum rate depends on the cylinder volume of the buret used. Formula input possible.
Cylinder volume 2 mL
Range Default value Selection Range Default value Selection Range Default value Selection Range Default value Selection Range Default value Selection
0.01 ... 6.00 mL/min 1.00 mL/min maximum
Cylinder volume 5 mL
0.01 ... 16.00 mL/min 1.00 mL/min maximum
Cylinder volume 10 mL
0.01 ... 33.00 mL/min 1.00 mL/min maximum
Cylinder volume 20 mL
0.01 ... 66.00 mL/min 1.00 mL/min maximum
Cylinder volume 50 mL
0.01 ... 166.00 mL/min 1.00 mL/min maximum
Comment
Optional comment on the time program command.
Input
64 characters
Generated variables
The following type 'TP.command number.Variable name' variables are generated by the command: Variable name Description CONC Data type Concentration of the dosed solution as long as one Number DVOL has been assigned to the Dosino. Dosed volume [mL] Number Aspirate
Dialog window: Method ? Time program ? Edit ? 'Device' - Aspirate
The device-dependent time program command Aspirate (command with feedback) draws in the specified volume via the defined port. As with the command Dosing, the cylinder is not filled automatically beforehand or afterwards. It should be possible to achieve the volume to be aspirated with a single piston stroke.
Parameters
Port
Port via which aspiration takes place. Formula input possible.
Range Default value Volume
1 ... 4 1
Volume to be transported. Formula input possible.
Range Default value Rate
0.0000 ... 50.0000 mL 1.0000 mL
The maximum rate depends on the cylinder volume of the buret used. Formula input possible.
Cylinder volume 2 mL
Range Selection Default value Range Selection Default value
0.01 ... 6.00 mL/min maximum maximum
Cylinder volume 5 mL
0.01 ... 16.00 mL/min maximum maximum