lcms
1Maintenance
To clean the electrospray spray chamber weekly
10Dampen a clean cloth and wipe the end of the capillary cap.
11Reinstall the spray shield.
12Use abrasive paper to gently clean the spray shield.
13Dampen a clean cloth and wipe the spray shield.
14Rinse the area around the spray shield.Figure 4
Rinsing the area around the spray shield
CAUTIONDo not spray directly toward the tip of the capillary. This can cause pressure surges in
the vacuum system.
15Wipe the area around the spray shield.
16Close the spray chamber.
17Reinstall the electrospray nebulizer.