lcms
Maintenance
To remove the electrospray nebulizer1
To remove the electrospray nebulizer
When required
Tools required
Parts requiredWhen you need to access the nebulizer for maintenance Gloves, clean None NOTERecent residue should be soluble in the mobile phase. If you are not sure what mobile
phase was used recently, a mixture of 50% isopropanol and 50% water works well as
ageneral cleaning solution.
Figure 5
Electrospray nebulizer
1Shut off the flow of LC solvent.
2Shut off the flow of nebulizing gas.3Slide back the plastic cover from over the nebulizer.
4Disconnect the LC tubing and nebulizing gas tubing from the nebulizer.
5Turn the nebulizer counterclockwise until it disengages from the retaining
screws.
6Gently lift the nebulizer out of the spray chamber.
WARNINGThe tip of the nebulizer may be very hot. Allow it to cool before handling it.