Agilent三重四极杆LCMS维护手册(17)

2021-02-21 08:07

lcms

Maintenance

To adjust the electrospray nebulizer needle1

To adjust the electrospray nebulizer needle

When requiredAfter replacing the electrospray nebulizer needle or if symptoms indicate the

needle is not correctly adjusted

Tools required Adjustment fixture (p/n G1946-20215)

Gloves, clean

Magnifier (p/n G1946-80049)

Wrench 3-mm, open-end (p/n 8710-2699)

Wrench ¼-inch x 5/16-inch, open-end (p/n 8710-0510)

Parts requiredNone

1Install the nebulizer in the adjustment fixture. Figure 9

2Loosen the needle holder locknut.

3Position the magnifier so you can see the tip of the nebulizer.

4Adjust the needle holder until the needle is even with the tip of the

nebulizer.