lcms
Maintenance
To clean the electrospray spray chamber daily1
Figure 1Opened electrospray spray chamber
5Rinse the interior of the spray chamber with the current mobile phase or
with a mixture of isopropanol and water.
WARNINGSome mobile phases are dangerous. Use the degree of caution appropriate for the
mobile phase being used.
6Wipe the interior of the spray chamber with a clean, lint-free cloth.
7Rinse the area around the spray shield.
CAUTIONDo not spray directly toward the tip of the capillary. This can cause pressure surges in
the vacuum system.
8Dampen a clean cloth with the mobile phase.
9Wipe the spray shield and the area around the spray shield.
10Close the spray chamber.
11Reinstall the electrospray nebulizer.
WARNINGUse the weekly cleaning procedure if symptoms of contamination persist or if the spray
shield or capillary cap show significant discoloration that cannot be removed by the regular,
daily cleaning.