lcms
1Maintenance
To clean the electrospray spray chamber daily
To clean the electrospray spray chamber daily
When required
Tools requiredDaily or at the end of each shift or anytime you suspect carryover contamination from one sample or analysis to another Cloths, clean, lint-free (p/n 05980-60051)
Gloves, clean
Isopropanol, reagent grade or better
Mobile phase, current
Wash bottle, clean
Water, reagent-grade or better
Parts requiredNone
Recent residue should be soluble in the mobile phase. If you are not sure what mobile
phase was used recently, a mixture of 50% isopropanol and 50% water works well as a general cleaning solution.NOTE
1Prepare the mobile phase you have been using.
2Turn off the spray chamber.
WARNINGThe electrospray spray chamber operates at high temperatures. Allow sufficient time to cool down before cleaning.
3Remove the electrospray nebulizer.
4Open the spray chamber (Figure1).