Agilent三重四极杆LCMS维护手册(8)

2021-02-21 08:07

lcms

1Maintenance

To clean the electrospray spray chamber daily

To clean the electrospray spray chamber daily

When required

Tools requiredDaily or at the end of each shift or anytime you suspect carryover contamination from one sample or analysis to another Cloths, clean, lint-free (p/n 05980-60051)

Gloves, clean

Isopropanol, reagent grade or better

Mobile phase, current

Wash bottle, clean

Water, reagent-grade or better

Parts requiredNone

Recent residue should be soluble in the mobile phase. If you are not sure what mobile

phase was used recently, a mixture of 50% isopropanol and 50% water works well as a general cleaning solution.NOTE

1Prepare the mobile phase you have been using.

2Turn off the spray chamber.

WARNINGThe electrospray spray chamber operates at high temperatures. Allow sufficient time to cool down before cleaning.

3Remove the electrospray nebulizer.

4Open the spray chamber (Figure1).