Agilent三重四极杆LCMS维护手册(7)

2021-02-21 08:07

lcms

Maintenance

To flush the nebulizer1

To flush the nebulizer

When required

Tools required

Parts requiredDaily or at the end of each shift (or more often) to flush traces of samples and buffers out of the tubing, valves and nebulizer Acetonitrile, HPLC-grade or better Water, HPLC-grade or better None

1Make sure acetonitrile and water are two of the solvents installed in your

liquid chromatograph.

NOTEThis procedure applies to both electrospray and APCI nebulizer.

2Set the liquid chromatograph to pump a mixture of 90% acetonitrile and

10% water at 2 ml/minute.

3Pump this mixture through the nebulizer for 3 minutes.

NOTEThis is a good general-purpose flushing mixture but you may need to adjust it based on the

solvents, samples and buffers you are using. For example, a mixture of 50% acetonitrile and 50% water works well for removing salts.