To replace the electrospray nebulizer needleFigure 7

Sliding the non-tapered end of the needle through the new needle holder7Push a new ferrule, flat-side first, onto the needle.

8Be sure the needle does not extend from the ferrule.

9Reinstall the locknut and the union. Hand tighten the union.

10Hold the needle holder steady with a 3-mm wrench. Tighten the union

one-quarter to one-half turn to compress the ferrule.

Figure 8Tightening the union screw to compress the ferrule

11Tighten the locknut against the union.

12Pull carefully on the needle to ensure the needle is held firmly in place.

13Replace locknut and washer.

14Insert the needle into the nebulizer shaft.