1Maintenance

To adjust the electrospray nebulizer needleFigure 10

Adjusting the needle holder

5Tighten the locknut. Make sure this does not change the position of the

needle.

6Remove the nebulizer from the adjustment fixture and reinstall it in the

electrospray spray chamber.

7Be very careful not to hit the tip of the nebulizer against anything. Any

damage will have a large, negative effect on system performance.

NOTERecord this procedure in the Maintenance Logbook.